Three Phase CCD Clocking
Charge transfer through CCD shift registers occurs after integration to relocate accumulated charge information to the sense amplifier, which is physically separated from the parallel pixel array. Several clocking schemes, including the three phase technique illustrated in Figure 1, are utilized to transfer charge from the collection gates to the output node.
Three phase CCD clocking improves spatial resolution over that obtained in four phase devices, yet requires only three gates per pixel. This scheme differs from four phase clocking by using only one storage gate and two barrier gates, which allows for faster frame rates and the fabrication of higher density and resolution CCDs. Another advantage of the technique is the formation of three independent polysilicon layers, one for each set of gate electrodes, so that clock phases are not forced to operate on two gates embedded in the same polysilicon layer. This innovation increases device fabrication yield.
Three-Phase CCD Clocking Scheme
Explore how charge transfer occurs from the shift registers to the output node in a three-phase charge-coupled device clocking scheme.
A three phase CCD incorporates three individual polysilicon gate electrodes in each pixel cell, each of which requires a separate input clock signal to properly transport accumulated charge in a scheme that requires a more complex clocking arrangement than that described for the four phase CCD. The shift register illustrated in Figure 1 includes two complete pixel elements, for a total of six gates aligned along a common axis to form a column. The nature of electrostatic forces in the silicon substrate beneath the gates is determined by the voltage level applied to a particular gate by the clock input signal. High level voltages induce the formation of a potential "well" beneath the gate, whereas low level voltages form a potential barrier to electron movement.
The initial state at t(1) of the CCD illustrated in Figure 1 is gates P(1) and P(2) of both Pixels 1 and 2 having low-level voltages (forming potential barriers) and gates P(3) of the pixels with high-level voltages, forming potential wells filled with purple spheres that represent integrated charge (electrons). At t(2), gates P(1) at both pixels are brought to a high voltage, followed shortly, but not simultaneously, by gates P(3) assuming low voltages. The integrated charge now resides under gates P(1) of both pixels. In a similar manner, the charge can be further shifted at t(3) by changing the voltage applied to gates P (1) and P(2), to complete the cycle. At the start of the next cycle (t(4)), the voltage to gates P(2) shifts from high to low, followed by the voltage to gates P(3) shifting from low to high. The net result is that gates P(3) from both pixels now contain charge originally integrated at the previous pixel.