How to Check the Imaging Software Magnification Settings
Two software products can be used for length measurement:
- cellSens software (for PC)
- Stand-alone camera connection kit DP2-SAL/DP2-AOU control software
To confirm that the measurement settings are correct, youʼll need a 0.01 μm stage micrometer from Evident or a third party. These can be purchased from Evident.
Evident: Stage micrometer
Model: OB-M1/100micro
How to check the magnification:
Step 1: Set the objective micrometer on the microscopeʼs stage.
Step 2: Move the stage so that the micrometerʼs scales are in the center of the screen. Make sure theyʼre in focus.
*Image shown is from cellSens software
cellSens standard layout
cellSens simple layout
Stand-alone camera software (DP2-SAL)
Stand-alone camera software (DP2-AOU)
Step 4: Measure using the scale function and check that the results are correct.
✓ 100x objective: 50 µm (5 divisions)
✓ Objectives other than 100x: 100 µm (10 divisions)
cellSens Software Measurement Method
Note: For details about the stand-alone connection kit (DP2-SAL/DP2-AOU) measurement method, please refer to the softwareʼs instruction manual.
Step 5: Repeat step 4 for the objectives you are going to use.
*When you change objectives, be sure to verify the settings (see step 3).
If the measurement results differ from what you expect, thereʼs a possibility that the magnification setting is not correctly specified. Please check the magnification setting and change it if itʼs not correct.