미러 유닛 광학계

형광 Filter Cubes

문의처

UIS2 형광(Fluorescence) 미러 유닛은 형광(Fluorescence) 색소의 최대 성능을 보장하며, 균일하고 높은 광 투과율을 만들어줍니다. 필터의 높은 효율은 효율은 약한 여기 광을 사용할 수 있으므로 광표백(photobleaching)과 표본의 손상을 최소화합니다.

BX3와 IX3, 그리고 APX100 용

Unit Name
Excitation Filter
Emission Filter
Dichromatic Mirror
Ultraviolet Excitations
U-FUW
340-390
420IF
410
U-FUN
360-370
420IF
410
U-FUNA
360-370
420-460
410
Blue Violet Excitations
U-FBVW
400-440
460IF
455
Blue Excitations
U-FBW
460-495
510IF
505
U-FBWA
460-495
510-550
505
U-FBN
470-495
510IF
505
U-FBNA
470-495
510-550
505
Green Excitations
U-FGW
530-550
575IF
570
U-FGWA
530-550
575-625
570
U-FGNA
540-550
575-625
570
Yellow Excitations
U-FYW
540-585
600IF
595
형광(Fluorescence) 단백질 변이체
U-FCFP
425-445CFP
460-510CFP
455CFP
U-FGFP
460-480GFP
495-540GFP
490GFP
U-FYFP
490-500YFP
515-560YFP
515YFP
U-FRFP
535-555HQ
570-625HQ
565HQ
U-FMCHE
565-585
600-690
595

IX3 용

Unit Name
Excitation Filter
Emission Filter
Dichromatic Mirror
Ultraviolet Excitations
IX3-FUWXL
340-390
420IF
410
Blue Violet Excitations
IX3-FBVWXL
400-440
460IF
455
Green Excitations
IX3-FGWXL
530-550
575IF
570
형광(Fluorescence) 단백질 변이체
IX3-FCFPXL
425-445CFP
460-510CFP
455CFP
IX3-FGFPXL
460-480GFP
495-540GFP
490GFP
IX3-FYFPXL
490-500YFP
515-560YFP
515YFP
IX3-FMCHEXL
565-585
600-690
595

IX2와 BX2, 그리고 MVX10 용

Unit Name
Excitation Filter
Emission Filter
Dichromatic Mirror
Ultraviolet Excitations
U-MWU2
330-385
420
400
U-MNU2
360-370
420
400
U-MNUA2
360-370
420-460
400
Violet Excitations
U-MNV2
400-410
455
455
Blue Violet Excitations
U-MWBV2
400-440
475
455
U-MNBV2
420-440
475
455
Blue Excitations
U-MSWB2
420-480
520IF
500
U-MWB2
460-490
520IF
500
U-MNB2
470-490
520IF
500
U-MWIB3
460-495
510IF
505
U-MWIBA3
460-495
510-550
505
U-MNIB3
470-495
510IF
505
U-MNIBA3
470-495
510-550
505
Green Excitations
U-MWG2
510-550
590
570
U-MNG2
530-550
590
570
U-MWIG3
530-550
575IF
570
U-MWIGA3
530-550
575-625
570
U-MNIGA3
540-550
575-625
570
Yellow Excitations
U-MWIY2
545-580
610IF
600
형광(Fluorescence) 단백질 변이체
U-MCFPHQ
425-445HQ
460-510HQ
450
U-MGFPHQ
460-480HQ
495-540HQ
485
U-MYFPHQ
490-500HQ
515-560HQ
505
U-MRFPHQ
535-555HQ
570-625HQ
565

선택 표

형광(Fluorescence) 단백질 변이체

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
AmCyan
458
489
U-FCFP, U-FBVW
U-MWBV2, U-MNBV2, U-MCFPHQ
Azami-Green
492
505
U-FBWA, U-FBNA, U-FGFP, U-FBW, U-FBN
U-MWIB3, U-MWIBA3, U-MNIB3, U-MNIBA3, U-MGFPHQ
DsRed2
563
582
U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ
EBFP
380
440
U-FUNA, U-FUW, U-FUN
U-MWU2, U-MNU2, U-MNUA2, U-MNV2
ECFP
433
475
U-FCFP, U-FBVW
U-MCFPHQ, U-MWBV2, U-MNBV2, U-MNV2
EGFP
488
507
U-FBWA, U-FBNA, U-FGFP, U-FBW, U-FBN
U-MGFPHQ, U-MNIB3, U-MWIBA3, U-MWIB3, U-MNIBA3
EYFP
513
527
U-FBWA, U-FBNA, U-FYFP, U-FBW, U-FBN
U-MYFPHQ, U-MWIBA3, U-MNIBA3, U-MWIBA3, U-MNIBA3, U-MNIB3, U-MNB2, U-MWB2, U-MWIB3
Fucci (mAG1)
492/551
505/565
U-FBNA, U-FGFP, U-FBWA, U-FBW, U-FBN
U-MNIBA3, U-MGFPHQ, U-MNIB3, U-MWIBA3, U-MWIB3
GFP (S65T)
488
510
U-FBWA, U-FBNA, U-FGFP, U-FBW, U-FBN
U-MGFPHQ, U-MWIBA3, U-MNIBA3, U-MWIB3, U-MNIB3
GFP (WT)
395, 475
510
U-FBWA, U-FBNA, U-FGFP, U-FBW, U-FBN
U-MWIBA3, U-MNIBA3, U-MGFPHQ, U-MWIB3, U-MNIB
HcRed
588
618
U-FYW
U-MWIY2
Midoriishi-Cyan
472
495
U-FBWA, U-FBW, U-FBNA, U-FBN, U-FCFP, U-FGFP
U-MWIBA3, U-MWIB3, U-MNIBA3, U-MNIB3, U-MCFPHQ, U-MGFPHQ
mCherry
580
610
U-FMCHE, U-FYW
U-MWIY2
mRFP
584
607
U-FMCHE, U-FYW
U-MWIY2
mStrawberry
574
596
U-FMCHE, U-FYW
U-MWIY2
RFP (DsRed)
558
583
U-FRFP, U-FGW, U-FGWA, U-FGNA
U-MRFPHQ, U-MWIG3, U-MWIGA3, U-MNIGA3
Venus
515
528
U-FYFP
U-MYFPHQ, U-MNB2, U-MNIB3, U-MNIBA3, U-MSWB2, U-MWB2, U-MWIB3, U-MWIBA3

광활성화(Photoactivation)

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Dronpa-Green
503
518
U-FBN, U-FBNA, U-FBW, U-FBWA, U-FYFP
U-MGFPHQ, U-MNB2, U-MNIB3, U-MNIBA3, U-MWB2, U-MWIB3, U-MWIBA3, U-MYFPHQ
PA-GFP
504
517
U-FGFP, U-FBWA, U-FBNA, U-FBW, U-FBN
U-MGFPHQ, U-MNIB3, U-MWIBA3, U-MWIB3, U-MNIBA3

광변환(Photoconversion)

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Kaede (Green/Red)
509/573
518/582
U-FBN, U-FBNA, U-FBW, U-FBWA, U-FGFP,
U-FYFP, U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIB3, U-MWIBA3, U-MGFPHQ, U-MYFPHQ, U-MNB2, U-MNIB3, U-MNIBA3,U-MWB2, U-MNIGA3, U-MRFPHQ, U-MWIG3, U-MWIGA3

Carbohydrate Modification, Protein Modification (Labeling)

단백질 변형(Protein modification)

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Alexa Fluor 350
346
442
U-FUNA, U-FUW, U-FUN
U-MNUA2, U-MWU2, U-MNU2
Alexa Fluor 405
402
421
U-FUNA, U-FUW, U-FUN
U-MNUA2, U-MWU2, U-MNU2
Alexa Fluor 430
434
539
U-FBVW
U-MWBV2, U-MNBV2, U-MNV2
Alexa Fluor 488
495
519
U-FBWA, U-FBNA, U-FBW, U-FBN
U-MWIBA3, U-MNIBA3, U-MWIB3, U-MNIB3,UMNB2, U-MWB2
Alexa Fluor 532
531
554
U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ
Alexa Fluor 546
556
573
U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ
Alexa Fluor 555
555
567
U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ
Alexa Fluor 568
578
603
U-FGW, U-FGWA, U-FGNA, U-FRFP, U-FYW, U-MCHE
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ, U-MWIY2
Alexa Fluor 594
590
617
U-FYW, U-FMCHE
U-MWIY2
Allophycocyanin (APC)
650
660
U-FYW
U-MWIY2
7-amino-4-methylcoumarin-3-acetic acid (AMCA)
349
448
U-FUNA, U-FUW, U-FUN
U-MNUA2, U-MWU2, U-MNU2
BODIPY FL
505
513
U-FBWA, U-FBW, U-FBNA, U-FBN, U-FGFP
U-MWIBA3, U-MWIB3, U-MNIBA3, U-MNIB3
Cascade Blue
376
425
U-FUNA, U-FUW, U-FUN
U-MNUA2, U-MWU2, U-MNU2
Fluorescein-isothiocyanate (FITC)
494
518
U-FBWA, U-FBW, U-FBNA, U-FBN, U-FGFP, U-FYFP
U-MWIBA3, U-MNIBA3, U-MWB2, U-MNB2, U-MWIB3, U-MNIB3, U-MGFPHQ, U-MYFPHQ
Oregon Green 488
496
524
U-FBWA, U-FBW, U-FBNA, U-FBN, U-FGFP, U-FYFP
U-MWIBA3, U-MNIBA3, U-MWB2, U-MNB2, U-MWIB3, U-MNIB3, U-MGFPHQ, U-MYFPHQ
Phycoerythrin B (PE-B)
545
576
U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ
Phycoerythrin R (PE-R)
490, 565
578
U-FGNA, U-FGW, U-FGWA, U-FRFP
U-MNIGA3, U-MWIG3, U-MWIGA3, U-MRFPHQ
Rhodamine B-isothiocyanate (RITC)
543
580
U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ
Texas Red
595
615
U-FYW, U-MCHE
U-MWIY2
Tetramethylrhodamine-isothyocyanate (TRITC)
541
572
U-FGW, U-FGWA, U-FGNA, U-FRFP
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MRFPHQ
1-Anilinonaphthalene-8-sulfonic acid (ANS)
365
460
U-FUW, U-FUN
U-MWU2, U-MNU2
CPM (7-Diethylamino-3-(4'-Maleimidylphenyl)-4-Methylcoumarin)
~384
470
U-FUW, U-FUN
U-MWU2, U-MNU2
Cy2
489
506
U-FBWA, U-FBNA, U-FBW, U-FBN
U-MWIBA3, U-MNIBA3, U-MWIB3, U-MNIB3
Cy3
(512); 550
570; (615)
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3
Cy3.5
581
594; (640)
U-FYW, U-FMCHE
U-MWIY2
Dans (1-Dimethyl Amino Naphaline 5 Sulphonic Acid)
340
525
U-FUW, U-FUN
U-MWU2, U-MNU2
Dansyl chloraide
360
520
U-FUW, U-FUN
U-MWU2, U-MNU2
Dansyl hydrazine
340
525
U-FUW, U-FUN
U-MWU2, U-MNU2
Eosin
525
545
U-FBWA, U-FBNA, U-FBW, U-FBN
U-MWIBA3, U-MNIBA3, U-MWIB3, U-MNIB3
Fluorescamine
390
465/475
U-FUW, U-FUN, U-FNUA
U-MWU2, U-MNU2, U-MNUA2, U-MNV2
FluorX
494
520
U-FBWA, U-FBNA, U-FBW, U-FBN
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MSWB2, U-MWB2, U-MNB2
Lucifer Yellow CH
425
528
U-FBVW, U-FBW, U-FBW, U-FBWA
U-MWBV2, U-MNBV2, U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MWB2, U-MNIB2, U-MSWB2
Monochlorobimane (mBCI)
394
490
U-FUW, U-FUN, U-FUNA
U-MNUA2, U-MWU2, U-MNU2, U-MNV2
o-phtalaldehyde (OPT)
340
455
U-FUNA, U-FUW, U-FUN
U-MNUA2, U-MWU2, U-MNU2
Resorufin
572
583
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWG2, U-MNG2
SITS (Primuline)
395-425
450
U-FUW, U-FUN, U-FUNA
U-MNUA2, U-MWU2, U-MNU2, U-MNV2
SITS (Stilbene Isothiosulphonic acid)
365
460
U-FUW, U-FUN, U-FUNA
U-MNUA2, U-MWU2, U-MNU2

Carbohydrate modification

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Acrifl avine
436
520
U-FBVW, U-FBW
U-MWBV2, U-MNBV2, U-MWB2, U-MSWB2

핵(Nuclei), 핵산(Nucleic Acid) 염색

핵(Nuclei), 핵산(Nucleic Acid)

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Acridine Orange
500
530/630-650
U-FBW, U-FBN, U-FBWA, U-FBNA, U-FYFP
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Acridine Yellow
470
550
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MWB2, U-MNIB2, U-MSWB2
Acrifl avine-Feulgen
455
515
U-FBVW, U-FBW, U-FBWA
U-MWBV2, U-MNBV2, U-MWIB3, U-MWIBA3, U-MWB2, U-MSWB2
Auramine O-Feulgen
460
550
U-FBW, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MWB2, U-MNIB2, U-MSWB2
7-Amino Actinomycin D
546
647
U-FYW, U-FGW, U-FMCHE
U-MWIG3, U-MWG2, U-MNG2, U-MWIY2
bis-Aminophenyl-oxadizole (BAO) -Feulgen
380
470
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
BOBO 1
462
481
U-FBVW
U-MWBV2, U-MNBV2, U-MCFPHQ
BO PRO 1
462
481
U-FBVW
U-MWBV2, U-MNBV2
Chromomycin A3
445
575
U-FBVW
U-MWBV2, U-MNBV2, U-MSWB2
4, 6-diamidino-2-phenyl-indole HCI (DAPI)
345
455
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Ethidium Bromide
493
620
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWG2, U-MNG2, U-SWG2
Hoechst 33258
345
478
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Hoechst 33342
343
483
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Mithramycin
445
575
U-FBVW
U-MWBV2, U-MNBV2, U-MSWB2
Olivomycin
430
545
U-FBVW
U-MWBV2, U-MNBV2
Pararosaniline-Feulgen
570
625
U-FYW, U-FMCHE, U-FGW, U-FGWA, U-FGNA
U-MWIY2, U-MWG2, U-MNG2, U-MWIG3, U-MWIGA3, U-MNIGA3
POPO 1
434
456
U-FBVW, U-FCFP
U-MNV2, U-MWBV2, U-MNBV2, U-MCFPHQ
Propidium Iodide (PI)
536
617
U-FGW, U-FGWA, U-FGNA, U-FYW
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWIY2
Pyronin Y
540
570
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3
Quinacrine Mustard
423
503
U-FBVW
U-MWBV2, U-MNBV2
SYTOX Green nucleic acid stain
504
523
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Thiazole Orange
510
530
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
TOTO 1
509
533
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
TOTO 3
642
661
U-FYW, U-FMCHE
U-MWIY2
TO PRO 3
642
661
U-FYW, U-FMCHE
U-MWIY2
YO PRO 1
491
509
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
YOYO 1
491
509
U-FBW, U-FBN, U-FBWA, U-FBNA, U-FGFP, U-FYFP
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3

세포생사판별

생세포 염색

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Acridine Orange
500
530/630-650
U-FBW, U-FBN, U-FBWA, U-FBNA, U-FYFP
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
BCECF indicator
482(low pH)
503(high pH)
520(low pH)
528 (high pH)
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Calcein
494
517
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Carboxyfl uorescein diacetate (CFDA)
496
515
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Fluorescein diacetate (FDA)
496
515
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3

죽은 세포 염색

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Ethidium Bromide
493
620
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWG2, U-MNG2, U-SWG2
Ethidium homodimer-1
528
617
U-FGW, U-FGWA, U-FGNA, U-FYW
U-MWG2, U-MNG2, U-MWIG3, U-MWIGA3, U-MNIGA3
Propidium Iodide (PI)
536
617
U-FGW, U-FGWA, U-FGNA, U-FYW
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWIY2

막(Membrane), 세포 기관(Organelle), 막 전위(Membrane Potential)

막

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
8-Anilino-1-naphthalene sulfonic acid (1,8-ANS)
385
485
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Dil
550
565
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3
DiO
484
501
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MSWB, U-MWB2, U-MNB2
Diphenylhexatriene (DPH)
351
430
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Filipin
350
480
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Fluorescamine
390
465
U-FUW, U-FUN
U-MWU2, U-MNU2, U-MNV2
Merocyanine 540
500
580
U-FGW
U-MWG2, U-MNG2, U-MWIG3

세포 소기관

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Acridine Orange
500
530/630-650
U-FBW, U-FBN, U-FBWA, U-FBNA, U-FYFP
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
BODIPY FL Ceramide
503
512
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3,U-MNIB3, U-MWIBA3, U-MNIBA3
DiO C6(3)
484
500
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MSWB, U-MWB2, U-MNB2
ER Tracker Blue-White DPX
374
575
U-FUW, U-FUN
U-MWU2, U-MNU2, U-MNV2
JC-1
485
530
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MSWB, U-MWB2, U-MNB2
Lyso Tracker Red DND-99
577
590
U-FGW, U-FGWA, U-FGNA, U-FYW
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWIY2
Mito Tracker Green
490
516
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Mito Tracker Red CMXRos
579
599
U-FGW, U-FGWA, U-FGNA, U-FYW
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWIY2
Nile Red (phospholipid)
552
636
U-FGW, U-FGWA, U-FGNA, U-FYW
U-MWG2, U-MNG2, U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWIY2
Rhodamine 123
507
530
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Pyronin Y
540
570
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWG2, U-MNG2

막 전위(Membrane potential)

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Di-4-ANEPPS
497
710
U-FBW, U-FBN
U-MWIB3, U-MNIB3
Di-8-ANEPPS
465
635
U-FBW, U-FBN
U-MWIB3, U-MNIB3
FM 1-43
473
579
U-FBW, U-FBN
U-MWIB3, U-MNIB3
RH160
577
>610
U-FGW, U-FGWA, U-FGNA, U-FYW
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWIY2
RH414
490
>662
U-FBW, U-FBN
U-MWIB3, U-MNIB3
RH421
541
>590
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3
RH423
520-550
>610
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3
RH795
560
650
U-FGW, U-FYW
U-MWIG3, U-MWIY2

세균 염색, 뼈 염색

Bacteria

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Acridine Orange
500
530/630-650
U-FBW, U-FBN, U-FBWA, U-FBNA, U-FYFP
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Acridine Yellow
470
550
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3, U-MWB2, U-MNIB2, U-MSWB2
Auramine O
460
550
U-FBVW, U-FBW
U-MWBV2, U-MNBV2, U-MSWB2, U-MWB2, U-MWIB, U-MWIB3
Berberine sulfate
430
550
U-FBVW
U-MWBV2, U-MNBV2
Coriphosphine O
460
575
U-FBW
U-MWB2, U-MWIB3
CTC
430, 480
630
U-FBVW, U-FBW, U-FBN
U-MWBV2, U-MNBV2, U-MWIB3, U-MNIB3
Hoechst 33258
365
465
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2

Bone

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Acidic Fuchsin (Pararosaniline)
540
630
U-FGW, U-FGN
U-MWG2, U-MNG2, U-MWIG3
Calcein
495
520
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Calcein Blue
375
435
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Tetracycline
390
560
U-FUW, U-FBVW
U-MWU2, U-MNV2, U-MWBV2, U-MNBV2
Xylenol Orange
440, 570
610
U-FGW, U-FGN
U-MWG2, U-MNG2, U-MWIG3

이온 농도(Ion Concentrations)

Calcium

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
BTC
400/480
540
U-FBN, U-FBNA, U-FBW, U-FBWA
U-MNV, U-MNIB3, U-MNIBA3, U-MWIB3, U-MWIBA3
Calcium Green
505
532
U-FBN, U-FBNA, U-FBW, U-FBWA
U-MNIB3, U-MNIBA3, U-MWIB3, U-MWIBA3
Fluo 4
495
510
U-FBN, U-FBNA, U-FBW, U-FBWA
U-MNIB3, U-MNIBA3, U-MWIB3, U-MWIBA3
Oregon Green 488 BAPTA-1
494
523
U-FBN, U-FBNA, U-FBW, U-FBWA
U-MNIB3, U-MNIBA3, U-MWIB3, U-MWIBA3
Oregon Green 488 BAPTA-5N
494
521
U-FBN, U-FBNA, U-FBW, U-FBWA
U-MNIB3, U-MNIBA3, U-MWIB3, U-MWIBA3

pH

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
BCECF
440/490
530
U-FBVW, U-FBN, U-FBNA, U-FBW, U-FBWA
U-MWBV, U-MNIB3, U-MNIBA3, U-MWIB3, U-MWIBA3

Nerve cell tracer, 기타

Nerve cell tracer

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Evans Blue
470-540
680
U-FBW, U-FBN
U-MWIB3, U-MNIB3
Fast Blue
360
410
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
Granular Blue
375
410
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2
True Blue
373
404
U-FUW, U-FUN, U-FUNA
U-MWU2, U-MNU2, U-MNUA2

기타

Excitation Maximum (nm)
Emission Maximum (nm)
BX3/IX3/APX100 Mirror Unit 호환용
BX2WI/MVX10 Mirror Unit 호환용
Aniline blue
370
509
U-FUW, U-FUN
U-MWU2, U-MNU2
Alexa Fluor 488-Phalloidin
495
518
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Calcofl uor white
440
500-520
U-FUW, U-FUN
U-MWU2, U-MNU2, U-MNV2
Euchrysine (Acridine Orange)
500
530/630-650
U-FBW, U-FBN, U-FBWA, U-FBNA, U-FYFP
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
NanoOrange
485
590
U-FBW, U-FBN
U-MWIB3, U-MNIB3
NBD-Phallacidin
465
536
U-FBW, U-FBN, U-FBWA, U-FBNA
U-MWIB3, U-MNIB3, U-MWIBA3, U-MNIBA3
Rhodamine-Phalloidin
540
565
U-FGW, U-FGWA, U-FGNA
U-MWIG3, U-MWIGA3, U-MNIGA3, U-MWG2, U-MNG2
sulforhodamine 101
586
605
U-FYW, U-FMCHE
U-MWIY2
Thiofl avin T
430
485
U-FBVW
U-MWBV2, U-MNBV2