Development of a New Fucci(CA) Application: A Fluorescent Probe for Visualizing Cell Cycles

Fucci (fluorescent ubiquitination-based cell cycle indicator) is a genetically encoded set of two fluorescent sensors for live cell-cycle monitoring. Fucci (CA), newly developed in 2017, labels nuclei with either mCherry (red) or mVenus (blue) in a cell-cycle-dependent manner. Fucci (CA) yields abundant expression of mCherry throughout G1 phase with an abrupt decrease in the red fluorescence at the end of G1 phase. Fucci (CA) can be used to reliably detect a short G1 phase and distinguish S and G2 phases, which previously was difficult to achieve.

Fucci (SA)

Fucci (SA)

Fucci (CA)

Fucci (CA)

Figure1: Visualization of Cell-Cycle Progression by Fucci (SA) and Fucci (CA)

Low Phototoxicity, Time-Lapse Image Acquisition of Undifferentiated ES Cells

Undifferentiated embryonic stem (ES) cells rapidly proliferate and are very delicate. Phototoxicity during time-lapse imaging may damage ES cells and reduce their proliferation speed, making it difficult to perform time-lapse imaging of ES cells under physiologically accurate conditions. The FV3000 microscope enables low-phototoxic time-lapse imaging by using extremely low laser power due to the highly efficient lightpath and sensitive detection devices. These properties of the FV3000 enabled a research group to perform a time lapse imaging experiment spanning 57 hours, in which three normal cell cycles of rapidly dividing undifferentiated ES cells were completely covered.

Figure2: Time-lapse observation of mouse ES cells labeled with Fucci (CA) 2.1

Imaging conditions
Specimen: Mouse ES cells
Objective: Silicone immersion objective (UPLSAPO30XS)
Microscope: FLUOVIEW FV3000 System
Laser: 445 nm (AmCyan), 594 nm (mCherry)

Detection of a Short G1 Phase

Fucci(CA) fully labels the G1 phase with a strong red signal, which enables precise measurement of phase boundaries, even in rapidly proliferating undifferentiated ES cells. Using time-lapse observation at the single cell level, mouse ES cells (mESCs) were found to proliferate with a doubling time of about 11 hours and a G1 phase of only one hour.

Figure3: Temporal profiles of fluorescence intensities (F.I.) of single cell nuclei expressing Fucci (CA) 2.1

Figure3: Temporal profiles of fluorescence intensities (F.I.) of single cell nuclei expressing Fucci (CA) 2.1

Features of FV3000 enabled this experiment

Fully Spectral System Provides High Sensitivity

The FV3000 series employs Evident's TruSpectral detection technology that diffracts light via transmission through a Volume Phase Hologram unit. This technology enables much higher light throughput when compared to conventional spectral detection units with reflection-type gratings.

High Signal-to-Noise Ratio (S/N) Under Low Excitation Light

The high-sensitivity detector unit in the FV3000 has up to four channels of GaAsP photomultiplier tubes (PMTs) that capture weak fluorescent signals with a maximum quantum efficiency of 45%. Additionally, Peltier cooling in these detectors reduces background noise by 20%. Together, these features enable high signal-to-noise ratio imaging, even with low excitation light.

Acquiring Physiologically Accurate Cell Cycle Images: Comment by Dr. Asako Sakaue-Sawano

Undifferentiated mESCs divide and move in a three-dimensional (3D) cell culture space, requiring XYZT imaging to monitor their proliferation. During time-lapse imaging, repeated laser scanning can potentially induce cell-cycle alterations in undifferentiated mESCs due to their high susceptibility to phototoxicity. The FV3000 microscope system permitted Dr. Asako Sakaue-Sawano and her colleagues to perform gentle 4-dimensional (XYZT) imaging experiments to accurately characterize the full cell cycle with very short G1 phase of rapidly proliferating mouse ES cells at the single cell level.

Acknowledgments
This application note was prepared with the help of the following researchers:
Laboratory for Cell Function Dynamics, RIKEN Center for Brain Science

Dr. Masahiro Yo

Dr. Masahiro Yo
(Responsible for the imaging experiments)

Dr. Asako Sakaue-Sawano

Dr. Asako Sakaue-Sawano
(First author)

Dr. Atsushi Miyawaki

Dr. Atsushi Miyawaki
(Team Leader)

Reference
For more details on the studies mentioned in this application note, please refer to the following article:
A. Sakaue-Sawano, et al. “Genetically Encoded Tools for Optical Dissection of the Mammalian Cell Cycle.” Molecular Cell, volume 68, issue 3 (October 2017): pp. 626–640.e5.

FV5000

Confocal Laser Scanning Microscope

  • Extraordinary clarity, speed, and reliability driven by groundbreaking innovations
  • SilVIR™ detectors deliver photon-level quantitation, exceptional sensitivity, and ultra-high signal-to-noise
  • Unmatched dynamic range captures the full signal spectrum and prevents saturation
  • High-speed 2K resonant scanning and high-density 8K galvo scanning in one platform
  • FLUOVIEW Smart™ software simplifies operation with intuitive controls and AI-powered automation
  • TruResolution™ auto correction collar optimizes focus for over 20 objectives
  • Modular design supports up to 10 laser lines and future multiphoton upgrades
  • Laser Power Monitor (LPM) ensures stable illumination and reproducible results over time

Learn More

FV5000MPE

Multiphoton Laser Scanning Microscope

  • Compact fiber-pigtailed lasers enable deep, quantitative imaging in scattering tissue
  • One-, two-, or three-line simultaneous MPE laser excitation for millimeters deep imaging
  • SilVIR™, TruAI, and TruSight™ technologies deliver outstanding signal-to-noise and clarity
  • MPE-optimized objectives, TruResolution™ auto correction collar, and automated IR laser alignment maintain sharp focus
  • Available as an FV5000 system upgrade or a complete MPE system
  • Fully tunable laser configurations available for more advanced multiphoton applications

Learn More

FV4000

Confocal Laser Scanning Microscope

  • Game-changing dynamic range for imaging from the macro scale to subcellular structures
  • Multiplex up to six channels simultaneously with TruSpectral technology
  • Redesigned high-speed, high-resolution scanners for fixed and live cell imaging
  • Improved depth and photosensitivity with pioneering NIR capabilities and renowned optics
  • Peace of mind with the reliable, repeatable SilVIR detector
  • Industry leading * ten laser lines with a broader spectral range from 405 nm to 785 nm

*As of October 2023.

Learn More

Super Apochromat Objectives

UPLSAPO-S/UPLSAPO-W

These super apochromat objectives provide spherical and chromatic aberration compensation and high transmission from the visible to the near infrared. Using silicone oil or water immersion media, which have refractive indexes closely matching that of live cells, they achieve high-resolution imaging deep in living tissue.

  • Compensate for both spherical and chromatic aberrations and high transmission from the visible to the near-infrared region
  • Silicone oil or water immersion media help achieve high-resolution imaging deep in living tissue and reduce spherical aberration as their refractive indexes closely match that of live cells

Learn More