CCD Clocking Schemes
CCD Clocking Schemes - Java Tutorial
Charge transfer through CCD shift registers occurs after integration to relocate accumulated charge information to the sense amplifier, which is physically separated from the parallel pixel array. This tutorial explores several clocking schemes that are utilized to transfer charge from the collection gates to the output node.
The applet initializes with a deactivated four-phase clocking scheme loaded into the window. Start the applet by selecting the Auto button, which will run the charge transfer automatically. Electrons comprising the integrated charge are represented by a green pool that is transferred through energy minima from one gate to the next. To step through the sequence manually, either click the Manual button to halt automatic transfer, or click on the Next >> button. Each click on the button will advance the tutorial a single step. Use the radio buttons to toggle between two-, three- and four-phase CCD clocking schemes.
The charge transfer process is termed readout, and is controlled by a series of clocks that operate on all gates in the array, including the transfer gate between serial and parallel registers and the photodiode reset gates. This cascade of clocking schemes is used to operate the CCD in a controlled and efficient manner.