Application Notes
Gaining the Edge: The Design and Development of a Device Equipped to Create 3D Ski Edge Measurements
Executive Summary
Within any sport, the fine-tuning of equipment can play a significant role in the outcome of competition. In the world of skiing and snowboarding, the sharpness of one's skis or board can drastically affect performance. At the highest levels of competition, the precision of edge sharpening becomes even more critical as the difference between finish times becomes smaller and smaller and every variable is increasingly magnified. To create a new level of sharpening accuracy, and measure ski edge resistance to a variety of surfaces and conditions, the Worcester Polytechnic Institute Sports Engineering Lab has designed and prototyped a device capable of extremely precise 3D ski edge imaging and measurement.
The objectives of this device are to measure the sharpness of ski and snowboard edges in two fundamental ways: functional and geometric, advancing the understanding of how the geometry of the edge relates to its performance. These measures can then be used to evaluate edges and preparation methods, making them especially important for competition as well as ski manufacturing and general ski maintenance.
Methodology
The current methods of assessing the sharpness of ski edges are qualitative and subjective, relying on tactical sensation. With this new device, sharpness is determined in the lab by applying a predetermined load to material like ice or a snow simulation material,oriented at fixed angle to the edge, then measuring the load necessary to cause movement perpendicular and parallel to the edge (Fig. 1). The geometry is measured and the curvature is calculated as functions of scale and position over the edge. Correlations are determined by linear regression analyses of performance versus curvature as a function of scale. Ski edges are measured with a laser scanning confocal microscope (Evident LEXT4100 [Previously Known as Olympus OLS4100]). Curvatures are calculated using Heron's method (Gleason at al. 2013)1.
Basic Principle of Performance Measurements
Created to provide equipment manufacturers, skiing facilities, athletes, and coaches with the means to measure ski and snowboard edges with a new level of accuracy, the ski edge measuring device developed by the WPI Sports Engineering Lab allows users to measure resistance to skidding perpendicular to an edge as well as sliding along an edge. Through the use of the LEXT OLS4100, this device can be used to create 3D topographic ski edge measurements specifically direct measurement and characterization of edge curvature.
Figure 1: Basic principle of device
performance measurements
Ratio of tangential to normal force T/N
- Measure of sharpness
- Similar to friction coefficient (µ=T/N)
Can also be applied tangential to edge to measure influence of edge sharpness on sliding resistance of an edged ski
The Importance of Proper Edge Roughness
The sharpness of a ski or snowboard edge plays a critical role in its performance. If an edge is too dull, it can cause control issues in both racing and recreational skiing. If icy hard-surface conditions are present, which occur occasionally in parts of the Northeast United States, control issues are even further magnified by dull edges. In the world of competitive racing, edges are sharpened even further to provide the ability to cut at high speeds with skis positioned at a sharp angle. Of additional consideration is the fact that highly sharpened edges can cause injury and should be treated with proper care and consideration both in use and while being handled.
Skiing Relative to Machining
Figure 2: Skidding, made possible by highly sharpened ski edges
Ski Edge Curvature
In the development of their new ski edge imaging and measuring device, the measurement specialists at WPI Sports Engineering Lab, led by Dr. Christopher Brown, set out to create a new way of looking at ski edge curvature, which provides an extremely precise means of viewing and analyzing ski edge roughness. By combining a physical assembly (Fig. 3) designed to provide optimal ski edge viewing with a laser scanning confocal microscope system (Evident LEXT4100 [Previously Known as Olympus OLS4100] [Fig.4]),Dr. Brown and his team have pioneered a new way to measure ski edge sharpness by accurately characterizing ski edge curvature.
Figure 3: Assembly prototype at WPI manufacturing labs. Edge sharpness configuration tests resistance perpendicular to ski edge. Load frame can be rotated to test resistance parallel to ski edge.
Figure 4: Evident LEXT4100 (Previously Known as Olympus OLS4100) laser scanning confocal microscope.
Roughness Imaging & Analysis
This new ski edge measurement system presents users with the means to create and analyze 3D topographic images of ski edge curvature. To showcase the effectiveness of this system, the WPI Sports Engineering Lab created and analyzed the following images of ski edges created with a variety of sharpening techniques. Terminology is based on a standard ski edge (Fig. 5).
Figure 5: This diagram represents the basic layout of and terminology associated with a standard ski edge
Figure 6: Profile of ski edge with no modifications.
Conclusion
Based on these 3D topographic images, Dr. Brown and his team were able to solidify the following conclusions regarding the ski edge sharpening techniques highlighted here:
- The method of sharpening (alternating vs. standard) will determine the direction of the bur at the tip of the ski edge.
- This bur can be beneficial to a skier by providing traction on turns; however, it can be disadvantageous by creating friction on straight paths, thus slowing the skier down.
- A ski sharpened with just a file is smoother than an unsharpened ski.
- A ski sharpened with a stone is even smoother than a ski sharpened with a ile.
- A progression system of sharpening starting with a standard ile and moving down toward a specialized sharpening stone will greatly decrease the surface roughness of the ski edge and provide a smoother ride to the skier.
Summary
The level of curvature imaging and edge measurement allowed by this new device is made possible by the implementation of direct geometric 3D confocal microscopy, which allows the calculation of curvature versus position and scale, a new calculation method used exclusively by this patent-pending system. In the past, curvature characterization had been problematic, as curvature varies with scale and position. Through their work with this new ski edge measurement system, WPI Sports Engineering Lab is pioneering multi-scale position-specific curvature characterization. This new system has been designed and prototyped to measure ski edge resistance to both skidding perpendicular to an edge and sliding along an edge. A provisional patent has been filed.
1Gleason, M. A., Kordell, S., Lemoine, A., Brown, C. A. (2013) Profile curvatures by Heron’s formula as a function of scale and position on an edge rounded by mass finishing. 14th International Conference on Metrology and Properties of Engineering Surfaces, Taiwan 17-21 June 2013.
Related products
LEXT OLS5500
Hybrid 3D Optical Profilometer
- Traceable surface measurements from the nanometer to micrometer scale
- Laser scanning microscopy (LSM), white light interferometry (WLI), and focus variation microscopy (FVM) in one award-winning platform
- First 3D optical profilometer to offer guaranteed accuracy and repeatability* for both LSM and WLI measurements
- WLI mode delivers up to 40x faster measurement throughput than conventional LSM
- Exceptional precision across surfaces with in-house engineered optics
- Intuitive interface and smart automation streamline operation for users of all levels
- AI-enhanced and high-throughput workflows with PRECiV™ software integration
*Based on Evident’s internal research as of October 2025. The guaranteed accuracy and repeatability apply only if the device has been calibrated according to the manufacturer’s specifications and is in defect free condition. Calibration must be performed by an Evident technician or an Evident-authorized specialist.
LEXT OLS5100
The LEXT™ OLS5100 laser scanning microscope combines exceptional accuracy and optical performance with smart tools that make the system easy to use. The tasks of precisely measuring shape and surface roughness at the submicron level are fast and efficient, simplifying your workflow and delivering high-quality data you can trust.