End-On Photomultipliers
In the end-on photomultiplier tube design, photons impact an internal photocathode and transfer their energy to electrons, which then proceed through a chain of electron multipliers termed dynodes ending in the anode. Instructions for operation of the tutorial appear beneath the applet window.
We are sorry, but this Java tutorial is no longer available.
The tutorial initializes with a single photon impacting the photocathode and producing corresponding photoelectrons, which then move through the dynode chain. Use the Stop button to start and stop photon flow, and the Gain slider to control the number of photoelectrons in the tutorial. The photoelectrons flow through the dynode chain and eventually are captured by the anode. When the Stop button is activated, the photomultiplier drawing reverts to a new illustration that diagrams photoelectron flow through the photomultiplier tube.
The spectral response, quantum efficiency, sensitivity, and dark current of a photomultiplier tube are determined by the composition of the photocathode. The best photocathodes capable of responding to visible light are less than 30 percent quantum efficient, meaning that 70 percent of the photons impacting on the photocathode do not produce a photoelectron and are therefore not detected. Photocathode thickness is an important variable that must be monitored to ensure the proper response from absorbed photons. If the photocathode is too thick, more photons will be absorbed but fewer electrons will be emitted from the back surface. However, if the photocathode is too thin, too many photons will pass through without being absorbed. The photomultiplier used in this tutorial is a end-on design, which uses an opaque and relatively thick photocathode of precise thickness as well as composition. Photoelectrons are ejected from the front face of the photocathode and angled toward the first dynode. Due to differences in design, side-on photomultiplier tubes often demonstrate higher quantum efficiencies than end-on tubes having photocathodes of similar composition.
In microscopy, the incident illumination is dispersed over a substantial portion of the photocathode, although the sensitivity is rarely uniform over the entire surface. In side-on designs, the upper half of the photocathode is typically 20 to 30 percent more sensitive than the lower half. The number of dynodes present in photomultipliers also varies by design. End-on tubes can have a series of up to 14 electrodes, whereas the side-on designs are usually limited to no more than 9 dynodes. For this reason, side-on photomultipliers do not achieve the high electron gains exhibited by end-on tubes.