White Paper

Contact Us

New Imaging Possibilities Using the World's First Multi-Immersion Silicone Gel Objective

Combining the ease of dry objective lenses with the optical performance of immersion lenses.

Introduction

The objective lens is the key component that determines the optical performance of a microscope. Traditionally, two types of objectives have been available: dry lenses and immersion lenses. Immersion objectives are widely used in life science research due to their ability to capture high-resolution and bright images.

However, users often report several challenges associated with immersion objectives:

To address these challenges, Evident developed the world’s first multi-immersion silicone gel objective. Featuring the LUPLAPO25XS objective lens and SILGEL-L25 silicone gel pad, this breakthrough technology combines the ease of dry objectives with the optical performance of immersion lenses.

This white paper explores these technologies in detail, highlighting the advantages of this unique objective lens system compared to conventional lens types.

Objective Lens Types

Microscope objectives fall into three categories:

1. Dry objective lenses

Dry lenses operate with air between the objective lens and the sample, making them easy to use. They are suitable for observation at low magnification and for wide-area observation by stitching together multiple images captured at high magnification. They also enable smooth transitions between macro and micro observations.

2. Immersion objective lenses

These lenses employ a liquid medium—typically silicone oil or water. Immersion lenses can be designed with high numerical aperture (NA) optics that deliver a high signal-to-noise ratio (SNR) and high-resolution images.1 By using an immersion liquid that matches the refractive index of the biological sample, spherical aberrations can be reduced in deep observation, and bright and high-resolution images can be acquired.

3. Silicone gel objective lenses

Using a semi‑solid silicone gel pad rather than liquid, these lenses maintain optical clarity without the drawbacks of liquid handling. The gel is stable and unaffected by the external environment, so it maintains moisture for reliable performance even in long-term imaging experiments.

Table 1. Comparison of Objective Lens Types and Performance

Dry objective Immersion objective Silicone gel immersion objective
Optical performance High SNR observation
High-resolution observation
Reduced depth crosstalk
Operational Performance Macro ↔ micro switching
Wide-area stitching/multipoint observation
Image acquisition stability
○ Good performance; ◎ Very good performance; △ Has challenges.

Challenges of Using Conventional Immersion Objective Lenses

While offering high magnification and resolution, conventional immersion objective lenses present several challenges.

1. Complicated lens switching

When starting with a low-magnification dry lens to position the specimen and switching to an immersion lens for observation, several issues may occur: loss of the focus position, air bubbles in the immersion solution, or an insufficient immersion medium.

When these issues occur, users must temporarily halt observation and clean the container and objective lens. This wastes labor and time, and it may negatively impact the specimens. For example, removing specimens from the stage-top incubator may cause damage to them. For specimens floating in the culture medium, the cleaning process may cause the specimens to shift, which may affect the reproducibility of the experiment. This leads to a case where the dry lens cannot be used after using the immersion one (Figure 1).

Ideally, it should be possible to switch between immersion and dry lenses without changing the specimen position, just like when switching between dry lenses, enabling imaging to run smoothly without interrupting observation.

Figure 1. It is possible to switch from a dry lens (left) to an immersion lens (right), but not the other way around as liquid remains on the specimen after immersion use.

2. XY observation range constraints

With immersion objective lenses, when the stage is moved beyond a certain range in the XY direction during specimen observation, the immersion liquid fails to follow the objective lens, causing the liquid to run out and making widefield observation impossible (Figure 2).

Figure 2. The immersion liquid runs out when the stage moves to another position, causing an air gap between the objective lens and specimen.

In recent years, there has been a growing demand for efficient observation of a large number of spheroids and organoids in multi-well plates. However, the limited XY observation range of immersion lenses makes it difficult to smoothly observe multiple wells in a plate, and dry lenses are unsuitable when deep observation is required.

Ideally, the system should be able to freely perform imaging over a wide area and multipoint observation at any position without concern for XY observation range restrictions.

3. Unstable retention of the immersion solution during prolonged imaging

During extended imaging experiments, image acquisition may fail due to air bubbles or an insufficient immersion solution caused by the external environment (Figure 3). If this issue leads to re-experimentation or recreation of specimens, the efficiency of the research is greatly reduced.

Ideally, the system should be able to stabilize immersion retention and perform imaging for extended periods of time without being affected by the external environment.

Figure 3. Immersion liquid can run out during time-lapse imaging, leading to data loss.

To overcome these challenges with immersion objective lenses, we focused on innovating semi-solid gels instead of liquids. This led to the development of an objective lens using silicone gel as the immersion material.

Developing a Silicone Gel Immersion Objective Lens

There were three technical hurdles in designing an objective lens that uses a semi-solid gel as its immersion medium.

1. Developing the optimal gel substrate

The following properties are required for a gel to be suitable for microscopic imaging.

Since the gel is pressed against the specimen during imaging, it is important that the optical properties, such as refractive index and Abbe number, remain the same before and after compression.

In addition, a material with low autofluorescence is essential to obtain fluorescence images with good SNR. The material must be soft and easily deformable for deep observation, but hard enough to return to its original shape without leaving gel residue in the container when the gel is removed from the specimen. Through a trial-and-error process of selecting the gel material, including optimizing the formulation conditions and shape, we succeeded in developing Evident's patented silicone gel pad technology that satisfies both optical and mechanical properties (Figures 4 and 5, Table 2).

Figure 4. Dry (left) and silicone gel pad (right) objectives can be easily switched.

Figure 5. Container appearance after gel adhesion leaves no gel residue on the container, so the cleaning step is eliminated.

Table 2. Evident’s Silicone Gel Properties for Microscopic Imaging

Gel Properties
Advantages for Microscopic Imaging
Optical Properties
Optical performance equivalent to silicone oil
  • Refractive index, Abbe number, transmittance, and autofluorescence do not change before and after compression, enabling acquisition of high SNR images during fluorescence observation.
  • No birefringence, and DIC observation is supported.
Mechanical Properties
Flexible and durable gel material
  • Combined with LUPLAPO25XS, observation down to 1 mm depth is possible
  • The material can be used continuously for about one month, reproducing its original shape even after several thousand adhesions and removal operations.
  • Low adhesion of the material to the container enables lens switching without shifting the observation position or cleaning the container.
Stability/Safety
Maintains properties in safe materials
  • The material is resistant to moisture and heat. Its properties remain unchanged over a long period of time with no expiration date.
  • The gel shape and physical properties do not change, so imaging experiments can be performed more stably and over a longer period of time without being affected by the external environment.
  • Safe and reliable materials that are gentle to the human body and the environment are used.

Reference: SDS

2. Frictional properties of the gel and container

Although silicone gel is an excellent material with both optical and mechanical properties, a certain frictional force exists between the gel and specimen. Therefore, if the specimen is moved in the XY direction beyond a certain range, a portion of the gel separates from the cover glass and a portion of the light does not pass through the immersion medium, resulting in image quality degradation (Figure 6).

Figure 6. If part of the silicone gel detaches from the cover glass after XY movement, then some light cannot pass through the immersion medium, impacting image quality.

To solve this issue, the cellSens software of the IX85 automated inverted microscope system is equipped with a control function dedicated to silicone gel immersion observation.

The risk of image quality deterioration is minimized by using the following automated control method: the silicone gel pad lens is removed from the specimen during XY movement (escape), the XY movement is performed, and the silicone gel pad lens is placed in close contact with the specimen again (Figure 7).

Figure 7. The silicone gel pad moves with the objective lens during multipoint acquisition.

The silicone gel pad lens and this control method enable high-quality images with a wide field of view, high signal-to-noise observation, high-resolution observation, and depth observation. These capabilities are advantageous, for example, in the observation of spheroids and organoids cultured in multiple wells of a multi-well plate, which requires imaging multiple points and a wide area. This method enables screening with high image quality.

The software's stage navigator displays the range that can be observed with a silicone gel immersion objective without escaping Z (Figure 8). This makes it possible to use the stage while knowing the observable area, minimizing the risk of going outside the range and impacting the gel immersion medium. As a result, users can capture high-quality images with confidence.

Figure 8. Visualization of the area observable with a silicone gel objective.

Establishment of a High-Quality Gel Production Method

Imaging gels must be of high quality. Contamination of foreign matter or bubbles leads to images with low signal-to-noise ratios, while rough gel surfaces significantly reduce durability performance and increase wear frequency. However, since there was no existing gel suitable for imaging applications, it was necessary to establish a manufacturing method from scratch and build a stable production system.

To give a simplified overview of the gel manufacturing method, the silicone gel solution is poured into a mold, cured, and then removed from the mold to be manufactured. A particular challenge was the surface precision of the gel. It is not enough to make the surface of the mold high precision. Through repeated studies, selection of a mold material suitable for the gel material, and optimization of the material and mold temperature conditions, it became possible to manufacture gels with a high-precision surface comparable to the surface of a lens.

To eliminate foreign matter and air bubbles, a high-class cleanroom environment was established at a level comparable to that used in the manufacture of optical elements. In addition, by setting strict inspection standards, we have established a production line that ensures consistent quality. These manufacturing environments have enabled us to achieve mass production of a high-quality silicone gel with stable and consistent optical and mechanical properties.

Development of a Multi-Immersion Objective Lens

In addition to the development of silicone gel, we used the lens polishing and assembly techniques from our renowned X Line™ series to develop a compatible objective lens: the LUPLAPO25XS. This special multi-immersion lens can be used with both our silicone gel pad technology and liquid forms by changing the attachment on the tip of the objective lens (Figure 9).

Figure 9. Silicone gel immersion (left) and water immersion (right) objective lenses.

The LUPLAPO25XS currently has the widest observation range (actual field of view is approximately 1 mm) and longest working distance (WD: 2 mm) of our silicone immersion objective lenses (Figure 10). The lens can also correct the refractive index from water to silicone oil (refractive index: 1.33 to 1.40).

Figure 10. Working distance and magnification of Evident’s silicone immersion objective lenses.

The wide field of view, long working distance, and refractive index range required for specimens in culture enable clear observation of spheroids and organoids over a wide area and down to a depth of 1 mm. The image below is a comparison of a dry objective lens with a similar NA (Figure 11), and it can be confirmed that silicone gel immersion allows clear observation of spheroids to a depth of 1 mm.

Figure 11. XYZ image comparison. Left: LUPLAPO25XS silicone gel pad objective, NA 0.85. Right: UPLXAPO20X dry objective, NA 0.8. Cleared HeLa* 1 cell spheroid (cyan: nuclei, magenta: microtubules). Captured on the IXplore IX85 Spin spinning disk confocal microscope. *To learn more about the origin of HeLa cells, visit henriettalacksfoundation.org

Summary

By developing the world's first multi-immersion silicone gel objective lens, we have successfully solved the potential challenges of using immersion objective lenses. This innovative lens technology offers several key advantages:

By facilitating the stable acquisition of high-quality imaging data, the multi-immersion silicone gel objective lens will contribute to improving the quality and efficiency of life science research.

References

1. Samuelsson, Andrew. “Silicone Immersion Objectives Answer the Call for Higher Resolution.” EvidentScientific.com. Originally published on Photonics Spectra. Accessed July 1, 2025.

Author

Kengo Osawa

Micro-Imaging Solutions R&D, Imaging Optics

Evident, Japan

IXplore IX85

IXplore IX85

The IXplore™ IX85 platform delivers an unmatched level of customizability, allowing you to design or build an intelligent, high-performance imaging system that meets your specific goals. And with an industry-leading field number (FN) of 26.5mm plus an array of advanced end-to-end imaging and workflow features, the IXplore™ IX85 means you can capture more than ever before while dramatically reducing your acquisition times.

Learn More

FV5000

FV5000

Confocal Laser Scanning Microscope

  • Extraordinary clarity, speed, and reliability driven by groundbreaking innovations
  • SilVIR™ detectors deliver photon-level quantitation, exceptional sensitivity, and ultra-high signal-to-noise
  • Unmatched dynamic range captures the full signal spectrum and prevents saturation
  • High-speed 2K resonant scanning and high-density 8K galvo scanning in one platform
  • FLUOVIEW Smart™ software simplifies operation with intuitive controls and AI-powered automation
  • TruResolution™ auto correction collar optimizes focus for over 20 objectives
  • Modular design supports up to 10 laser lines and future multiphoton upgrades
  • Laser Power Monitor (LPM) ensures stable illumination and reproducible results over time

Learn More

IXplore IX85 SpinXL

IXplore IX85 SpinXL

See more and discover faster with an unparalleled 26.5 mm field of view (FOV).

  • Capture more fast cellular dynamics in a single frame at imaging speeds up to 498 fps
  • Precise, gentle live-cell imaging maintains sample viability
  • Supports a wide range of applications and users with speed, precision, and flexibility
  • Streamlines tasks throughout the lab with intelligent automation and easy workflows
  • Evolves with your research via a wide wavelength range, tailored disk options, and expandable imaging modalities
  • Powered by CrestOptics spinning disk technology and enhanced with the IXplore IX85 platform

Learn More

IXplore IX85 SpinSR

IXplore IX85 SpinSR

Capture ultra-small cellular dynamics in stunning super-resolution detail.

  • Discover more with high-speed imaging in resolution down to 120 nm
  • Designed for prolonged cell viability in long-term time-lapse experiments
  • Powered by Yokogawa spinning disk technology
  • Enhanced with the IXplore IX85 platform and Evident’s TruSight SR super-resolution algorithm

Learn More